Spectral preprocessing and similarity matching (qualitative identification)

Preprocessing & Spectral Similarity — Qualitative Identification
Food Analysis Lab · 4-hour session
Raman/NIR preprocessing · building a spectral library · similarity metrics · NNLS mixture preview
ALS baseline (λ, p)SNV/MSCSG smoothing & derivativesPearson r / cosine / SAMLibrary matchingNNLS mixture fitting
Before we start, look at this packet of powder
A packet of white powder says “sugar substitute” —
is it aspartame? Sucralose? Erythritol? Or a blend of several?

To the eye, all four are fine white powders that look identical. But measure one spectrum and the shape never lies.

But don’t rush to match: measure the same powder again on another instrument, or packed a different way, and the raw spectrum can look completely different — a fluorescence background (baseline) buries the signal, different particle sizes cause scatter that tilts the baseline, and the intensity units of different instruments do not line up. Today we learn to remove this “noise” first, then use similarity metrics to say scientifically “which entry in the spectral library this powder resembles most”.

Today’s task: use the pure-compound spectra the whole class measured in W10 to build a small spectral library, then identify the TA’s blind samples and commercial sugar substitutes — and finally compare whose matching scores are the most convincing.

W11 / COURSE TOUR

Remove measurement effects, then identify spectra

Full narration and captions; simulated spectra, not measured fingerprints. Player and downloads · Interactive lab

W11 / PREDICT · OBSERVE · EXPLAIN

Six-act lab: choose preprocessing with evidence

Today’s learning goals

After W11 you should be able to…

  • Explain why raw spectra cannot be compared directly: three problems — fluorescence baseline, particle-size scatter, intensity differences
  • Apply cosmic-ray spike removal and ALS baseline removal to a Raman spectrum, and know how λ (smoothness) and p (asymmetry) affect the result
  • Apply SNV / MSC to remove scatter from an NIR spectrum, and use SG smoothing and 1st/2nd derivatives to sharpen detail
  • Compare spectra with four similarity metrics — Pearson r, cosine, SAM, Euclidean — and know whether each is sensitive to “intensity”
  • Build the class spectral library, identify the TA’s blind samples and commercial sugar substitutes, and report the top-3 rather than only the first place
  • Use NNLS for a first quantitative preview of an unknown mixture (Σcᵢ·referenceᵢ), leading into W16
  • In Orange OR11, read the distance matrix and dendrogram and find the library sample most similar to the unknown

Preprocessing tool: this lesson uses SpectraView (GitHub) ↗ (user guide ↗) for SG smoothing, baseline, SNV/MSC, derivatives, peak picking and .speclib library matching; as an easy start you can also use the Orange workflow OR11 ↗: File → Preprocess Spectra (ALS baseline subtraction → vector normalization) → Distances (Pearson) → Distance Matrix + Hierarchical Clustering, which only does qualitative identification (“which one is it most similar to”) — NNLS quantification of mixtures still needs Colab NB09.

What is a spectrum, really?

A spectrum = a string of numbers

Each wavelength/wavenumber is just one “column” and each intensity just one “cell” — the illustrative spectra in this lesson have 71 data points (one every 20 cm⁻¹), so one spectrum = 71 numbers in a row. Preprocessing (spike removal, baseline, SNV…) simply means “apply the same operation to every row”.

Raman shift (cm⁻¹)400420440460480
Intensity (a.u.)0.4310.4480.4571.3550.466

In Excel: one row = one spectrum, 71 columns = 71 wavenumbers; applying the same formula (e.g. SNV) across a whole row is “preprocessing”; 71 samples × 71 columns is the data table X used from W12 on.

Principle

Why can’t raw spectra be compared directly?

Why can’t raw spectra be compared directly? ① Fluorescence baseline Measured spectrum Fluorescence True signal The arched background lifts the true peaks and makes them less obvious ② Particle-size scatter ③ Intensity differences Same wavenumber, different intensity Same component, but particle size/packing changes both intercept and slope Measure two “identical” batches of sample on the same spectrometer and the numbers still don’t line up — that is why we preprocess first, then match.
Preprocessing ①

Cosmic-ray spike removal (despiking)

First use a “median filter” to estimate what each point should look like nearby, then see how far each point deviates from that estimate. The deviation is measured with a “robust standard deviation”, MAD (median absolute deviation), converted into a “modified z-score”; a deviation of more than 6 robust SDs is judged a spike and replaced with the median-filter value.

Position1234567
Raw value0.300.310.292.100.320.300.29
7-pt median filter0.300.300.300.300.300.290.29
Modified z-score0.00.7-0.7121.41.30.70.0
Corrected0.300.310.290.300.320.300.29

Point 4 is 2.10, z ≈ 121 (far above 6) → judged a cosmic-ray spike → replaced with the median-filter estimate 0.30. A window that is too wide may flag even real narrow peaks as spikes and delete them; too narrow may miss spikes — NB09/OR11 both use a 7-point window, z>6.

Preprocessing ②

ALS baseline: two parameters, λ and p

The defaults in NB09 / Orange OR11 are both λ=10⁶, p=0.01 — the common standard for this course. Orange path: Preprocess Spectra → ALS.

ALS baseline: if λ is too small, real peaks are subtracted as baseline too Measured spectrum (background + real peaks) λ large enough (e.g. 10⁶) = a stiff ruler: follows only the broad background and ignores the peaks λ too small = a soft rope: sinks into the real peaks and subtracts them as baseline p = 0.01: only 1% weight trusts points currently above the baseline — forcing the baseline to hug the bottom
Data · Raman

Raman preprocessing: spike removal + ALS baseline removal

Illustrative spectrum (not measured by this class), but the preprocessed curve is the result of actually running despike() + ALS(λ=1e6, p=0.01) on the raw curve at left, using the same formulas as NB09 clean_spectrum(). Before preprocessing there are 2 cosmic-ray spikes and a broad fluorescence baseline; if λ is too small, real peaks are eaten as baseline, and p controls how much “only points below the baseline are trusted”. The x-axis is evenly spaced Raman shift.
Preprocessing ③

SNV step by step

Formula: y′=(y−ȳ)/s, where ȳ = the mean of this spectrum itself and s = the standard deviation of this spectrum itself (population SD, dividing by n, not n−1).

Pt 1Pt 2Pt 3Pt 4Pt 5
y0.20.40.90.50.3
y′ (after SNV)-1.08-0.251.820.17-0.66

ȳ=0.46, s=0.242 (population SD). Replace each y with “how many standard deviations from the mean” and you get y′. Try replacing y with 2y+0.1 (intensity doubled, then shifted) and run SNV again: the result is exactly the same — that is why SNV is insensitive to both shift and scaling.
Excel: =(B2-AVERAGE($B2:$F2))/STDEV.P($B2:$F2) (note: use STDEV.P, not STDEV.S). MSC is conceptually the same; the difference is that it regresses against a reference spectrum (usually the mean spectrum), y=a+b·y_ref, then corrects with (y−a)/b.

Data · NIR

NIR preprocessing: SNV removes particle-size scatter

Before preprocessing: 3 spectra of the same component with different intercepts and slopes because of particle size/packing (illustrative data).
After SNV: each spectrum is re-standardized with its own mean and standard deviation, and the 3 spectra nearly coincide.

Intensity differences can also be handled by vector normalization (divide by vector length) or area normalization (divide by the area under the spectrum) — used in the next section.

Preprocessing ④

Savitzky–Golay smoothing and derivatives

Savitzky–Golay: fit a quadratic curve inside the window and keep only the fitted value at the “center point” Current window (15 points in practice) Output = this center point The window moves one step right and the quadratic is refit, repeating along the whole spectrum — the wider the window, the smoother, but the fatter the peaks; too low an order flattens the peak shape
Derivative intuition

Derivative = slope (think of differences)

1st derivative = the difference between neighboring points; 2nd derivative = the difference of differences.

Spectrum y1st derivative (difference)2nd derivative (difference again)
[1, 2, 4, 7][1, 2, 3][1, 1]

Add a constant to the whole spectrum (a uniform baseline shift, e.g. [6, 7, 9, 12]) and the 1st derivative is completely unchanged — differencing cancels the constant; a linear slope (the whole spectrum rising steadily in one direction) is also removed in the 2nd derivative. The cost: noise (random point-to-point jitter) changes quickly, and differencing amplifies it, so derivatives are often used together with smoothing (e.g. SG).

Data · NIR

NIR preprocessing: 1st/2nd derivative — peaks become negative minima

Illustrative data. The 2nd derivative turns into negative minima at the two overlapping peak positions (1450, 1500 nm), resolving the overlapping broad peaks more clearly; the cost is that point-to-point jitter (noise) is clearly amplified, so window width and polynomial order must balance “resolution” against “noise resistance”.
Preprocessing ⑤

Vector / area normalization — changes the intensity scale first

Mini-game ①8 preprocessing methods — which problem does each solve?0 / 8
Sort the 8 preprocessing methods by the main problem each solves: scatter / baseline / noise / intensity.
Similarity metrics

Four metric formulas for matching spectra

MetricFormulaSensitive to intensity magnitude?
Pearson rHow strongly the “shapes” of two spectra correlate (covered in W04 ↗)Not sensitive (unaffected by shift or scaling)
cosineTreat spectra as vectors; cosθ=(x·y)/(‖x‖‖y‖)Insensitive to scaling, but sensitive to shift
SAM (spectral angle)SAM = arccos(cosine) × 180/π (converted to “degrees”)Same as cosine — the smaller the angle, the more alike
Euclidean‖x−y‖, a direct numerical differenceVery sensitive — always normalize before matching
False-positive warning: matching a mixture often only hits the main component and looks like a single compound; and on raw NIR, two completely different compounds can still give r > 0.9 because they share broad O–H/C–H combination bands — in that case switch to derivative spectra or look at the Raman fingerprint region, and always report the top-3 scores rather than only the first place.
Worked example: ranking reversal

Without normalization, Euclidean picks the wrong answer

Unknown U=[0.10,0.35,0.90,0.40,0.15], candidate C=[0.12,0.32,0.95,0.38,0.22] (shape slightly different from U, but the absolute values are very close to U).

MetricU vs C
cosine0.9968
SAM4.59°
Pearson r0.9911
Euclidean (raw, not normalized)0.095
Euclidean (after vector normalization)0.080

Compare with the “truly correct” candidate: raw Euclidean d(U, truly correct)=0.510 is larger than the 0.095 of C above — raw distance wrongly judges C to be closer! After normalization, d(U, truly correct)=0.048 is smaller than C’s 0.080, so the judgment is correct. Pearson r, cosine and SAM are insensitive to intensity by design, so they never made this mistake. (Which candidate is truly correct? Work it out yourself in Calculation Challenge ⑥ later.)

Similarity metrics

Four metrics, four kinds of sensitivity

📈
Pearson correlation coefficient r
Looks only at the relative changes in “shape”; insensitive to both offset and scale — the most common way to check whether waveforms look alike
📐
Cosine similarity
The cosine of the angle between spectra treated as vectors; insensitive to scale but sensitive to overall offset
🧭
SAM spectral angle (radians/degrees)
The angle version of cosine — the smaller the angle, the closer the directions; SAM=0 means the two vectors point in exactly the same direction
📏
Euclidean distance
A direct numerical difference, very sensitive to intensity — always normalize before matching, or intensity differences will lead you astray
Data · spectral library

Library matching example: TA blind sample #1 (Raman, illustrative data)

Candidate compound ⇅Pearson r ⇅cosine ⇅SAM (°) ⇅Rank ⇅
Click a column header to sort. SAM is always converted from cosine (arccos(cos)×180/π), so the two columns agree. Preprocessing (spike removal + ALS + vector normalization) has been applied identically to the library and the unknown spectrum. There is a clear score gap between 1st place aspartame and 2nd place sucrose, so the call is safe; but if the top two scores are close (e.g. r differs by <0.02), always go back and confirm with a visual overlay and fingerprint-region peak positions — never rely on the numbers alone.
Mini-game ②Order the spectral identification workflow— / 6
Use ▲▼ to put the 6 steps of “identifying an unknown sample with a spectral library” in the correct order.
Mini-game ③For this identification task, Raman or NIR?0 / 6
Sort the 6 scenarios into “suits Raman” or “suits NIR”. Isomers = compounds with the same molecular formula but different atom arrangement (structure).
Orange starter path

OR11: distance matrix and dendrogram

Orange has no dedicated “Similarity / Library / Mixture” widgets — the OR11 identification is built from these four general-purpose widgets; NNLS mixture quantification and peak assignment can’t be done in Orange and need Colab NB09.
Data · mixture

Mixture preview: NNLS non-negative least-squares fitting

Illustrative data. Unknown spectrum ≈ 0.62 × sucralose + 0.34 × maltodextrin + intercept (proportions taken from NB09’s mix_demo); NNLS fits c₁=0.61, c₂=0.34, intercept=0.018, R²=0.995. “Least squares” = find the coefficients that minimize the “sum of squared residuals”; R²=1−Σresidual²/Σ(y−ȳ)² means the same as the “proportion of reference-value variance explained by the model” taught in W07 ↗. NNLS forbids negative coefficients, matching the physical fact that “a component’s content cannot be negative”; the reference spectra here are not vector-normalized, otherwise the linear “coefficient = proportion” relationship would be broken — a warm-up for W16 quantification.
Mini-game ④Preprocessing concepts: live quiz0 / 5
Peak assignment practice

Back to library matching: are these peak assignments correct?

Peak assignment practice (all approximate values; confirm with this class’s spectra) 40060080010001200140016001800 Raman shift (cm⁻¹) C–Cl stretch (sucralose) about 600–850 cm⁻¹ about 1003 cm⁻¹ Benzene ring breathing (aspartame) The numbers in the table are only a starting point — you must go back and compare with the peak positions “your own class measured” to complete the assignment.
The aspartame benzene-ring breathing mode falls near about 1003 cm⁻¹; the sucralose C–Cl stretch falls in the range about 600–850 cm⁻¹. These numbers are approximate literature values — always confirm where the peaks really fall using this class’s own W10 spectra, then use the same positions to assign today’s blind samples and commercial sugar substitutes.
Mini-game ⑤Decision challenge: how to interpret a match result?0 / 5
4-hour lab schedule

Lecture → preprocessing practice → build library → match → report

TimeContents
0:00–0:50Lecture: the three “can’t compare directly” problems, spike removal, ALS baseline, SNV/MSC/SG, four similarity metrics
0:50–1:40SpectraView preprocessing practice: try different parameters (λ, p, SG window/order) on the same spectrum and record your reasons for each choice
1:40–2:35Build the class spectral library from the W10 pure-compound spectra (or use Orange OR11) and match the TA’s blind samples
2:35–3:30Measure commercial sugar substitutes (Raman 785 nm + each group’s NIR) and match against the library
3:30–4:00Each group reports its top-3 results and similarity scores
🔦
Laser safety
With the Raman 785 nm/532 nm laser, never look into the beam path and never aim at anyone’s eyes; before adjusting, make sure the shield cover is closed
😷
Powder handling
Weigh sugar-substitute powder in a draft-free spot and avoid inhaling dust; clean the balance and bench after use
Mini-game ⑥Calculation challenge: SAM spectral angle (degrees)Unanswered
Two 5-point illustrative spectra: U = [0.10, 0.35, 0.90, 0.40, 0.15], C1 = [0.20, 0.55, 1.30, 0.60, 0.25]. First compute cosine = (U·C1)/(‖U‖‖C1‖), then SAM = arccos(cosine), and finally convert to degrees (°).
degrees
Homework

Colab NB09

📚
Library matching table
Include: the unknown sample, top-3 candidates, each similarity score (r/cosine/SAM), and the preprocessing methods used
🖼️
Overlay figure
Attach a screenshot overlaying the unknown spectrum and the top-1 candidate on one plot
🔍
One false-positive / ambiguous-match discussion
Pick a case where the top-1 and top-2 scores are close or the call was clearly wrong; explain the likely cause and how to confirm
🧮
NNLS practice
Pick any one mixture sample, run one NNLS fit, and record the coefficients and R²

Colab NB09: Open NB09 in Colab ↗ · Easy start (no coding): Orange workflow OR11 ↗. The report must include: a library matching table (top-3), the overlay, a false-positive/ambiguous-match discussion, and your reasons for the preprocessing parameters.

Self-check

By the end of today you should be able to…

Tap to tick.